The H. Pylori Ag Rapid Test Strip (Feces) is a rapid chromatographic immunoassay for the qualitative detection of antigens to H. Pylori in feces to aid in the diagnosis of H. Pylori infection.


Product Detail

Product Tags

Introducing the H Pylori Ag Rapid Test Device/Strip, an innovative product from QL biotech– a trusted establishment in the biotechnological market. These H Pylori Test Strips are designed to deliver quick, accurate, and dependable results. The fundamental principle of these testing strips revolves around the detection of H Pylori antigen in human stool specimens, providing an easy to read result in a short span of time. Perfect for clinics, hospitals, and even for personal use, these H Pylori Test Strips decrease the time consumed in conventional testing methods and increase the convenience for the user. QL biotech is known for its commitment to delivering high-quality products, and these test strips are no exception. They are created under strict quality control measures, ensuring you get precise results every time you use them.

PRINCIPLE

The H. Pylori Ag Rapid Test Devic/Strip (Feces) is a non‐invasive lateral flow assay, rapid, precise and easy to perform.This test makes use of specific antibodies against H. Pylori antigen adsorbed onto a reactive membrane. If H. Pylori is present in stool specimen, the specific antigen is bound by the second antibody which is conjugated with colloidal gold particles. A generic antibody, fixed onto the reactive membrane, in shape of the band, is able to capture the second conjugated antibody, assuring the correctness of the test performance.


Product Detail

Brand: QL

Specimens: : Feces

Reading time:10 minutes.

Pack:25 T

STORAGE: 2‐30°C

KIT COMPONENTS(Device)

Individually packed Test Devices/Strips

Each test contains a strip with colored conjugates and reactive reagents pre-spreaded at the

corresponding regions

Tubes with buffer

Phosphate buffered saline and preservative, extract the samples

Package insert

For operation instruction

PROCEDURE

    • Bring tests, specimens, buffer and/or controls to room temperature (15-30°C) before use. 
      1. Specimen collection and pre-treatment: 
      1) Unscrew and remove the dilution tube applicator. Be careful not to spill or spatter solution from the tube. Collect specimens by inserting the applicator stick into at least 3 different sites of the feces. 
      2) Place the applicator back into the tube and screw the cap tightly. Be careful not to break the tip of the dilution tube. 
      3) Shake the specimen collection tube vigorously to mix the specimen and the extraction buffer. Specimens prepared in the specimen collection tube may be stored for 6 months at -20°C if 
      not tested within 1 hour after preparation. 
      2. Testing 
      1) Remove the test from its sealed pouch, and place it on a clean, level surface. Label the test with patient or control identification. To obtain a best result, the assay should be performed within one hour. 
      2) Using a piece of tissue paper, remove the tip of the dilution tube. Hold the tube vertically and dispense 3 drops of solution into the sample well of the Test Device. 
      Avoid trapping air bubbles onto the sample pad, and do not drop any solution in observation window. 
      As the test begins to work, you will see color move across the membrane. 
      3. Wait for the colored band(s) to appear. The result should be read at 10 minutes. Do not interpret the result after 20 minutes.

    INTERPRETATION OF RESULTS


    • POSITIVE RESULT: 
      A colored band appears in the control band region (C) and another colored band appears in the T band region 
      NEGATIVE RESULT: 
      One colored band appears in the control band region (C). No band appears in the test band region (T) 
      INVALID RESULT: 
      Control band fails to appear. Results from any test which has not produced a control band at the specified reading time must be discarded. Please review the procedure and repeat with a new test. If the problem persists, discontinue using the kit immediately and contact your local distributor.
      NOTE: 
      1. The intensity of the color in test region (T) may vary depending on the concentration of aimed substances present in the specimen. Therefore, any shade of color in the test region should be considered positive. Besides, the substances level can not be determined by this qualitative test. 
      2. Insufficient specimen volume, incorrect operation procedure, or performing expired tests are the most likely reasons for control band failure.




    Understanding the importance of accurate H Pylori detection, we have designed these test strips with a high sensitivity rate. They are user-friendly and require minimal effort to obtain a clear result. Our H Pylori Test Strips are a great blend of modern technology and convenience, making the testing process much smoother and quicker. Choose QL biotech's H Pylori Test Strips for a hassle-free, rapid, and reliable detection process. Leverage the power of advanced technology for unrivalled accuracy and efficiency. Start your journey towards a better, quicker, and more precise testing process today.