Quick Norovirus Detection - Test One Step Dengue Test Kits

Short Description:

The Norovirus (GⅠ/GⅡ) Rapid Test Device (Feces) is a rapid visual immunoassay for the qualitative presumptive detection of Norovirus in human fecal specimens. This kit is intended to be used as an aid in the diagnosis of Norovirus infection.


Product Detail

Product Tags

Discover unparalleled accuracy and efficiency with QL Biotech’s revolutionary product - the Norovirus (GI/GII) Rapid Test Device, now encompassing the reliability of Test One Step Dengue Test Kits. In a world where speed and precision are paramount, this cutting-edge tool stands as a beacon for healthcare professionals and laboratories, offering a swift, one-step solution for the detection of Norovirus GI and GII types.

PRINCIPLE

The Norovirus (GⅠ/GⅡ) Rapid Test Device (Feces) has been designed to detect norovirus GⅠand GⅡ through visual interpretation of color development in the internal strip. The membrane was immobilized with anti-norovirus GⅠ antibodies and anti-norovirus GⅡ antibodies on the test region. During the test, the specimen is allowed to react with colored anti-norovirus GⅠ

antibodies colloidal gold conjugates and antinorovirus GⅡ antibodies colloidal gold conjugates, which were precoated on the sample pad of the test. The mixture then moves on the membrane by a capillary action, and interact with reagents on the membrane. If there were enough norovirus GⅠin specimens, a colored band will form at the T1 region of the membrane.

Similarily, If there were enough norovirus GⅡ in specimens, a colored band will form at the T2 region of the membrane. Presence of colored band(s) indicates a positive result, while its absence indicates a negative result. Appearance of a colored band at the control region serves as a procedural control. This indicates that proper volume of specimen has been added and membrane wicking has occurred.


Product Detail

  • Brand: QL

    Specimens: : Feces

    Reading time:10 minutes.

    Pack:20 T

    STORAGE: 2‐30°C

    KIT COMPONENTS

    Individually packed test devices

    Specimens dilution tube with buffer

    Disposable pipettes 

    Package insert

 

PROCEDURE

    • Bring tests, specimens, buffer and/or controls to room temperature (15-30°C) before use. 
      1. Specimen collection and pre-treatment: 
      1) Use the specimen collection container for specimen collection. Best results will be obtained if the assay is performed within 6 hours after collection. 
      2) For solid specimens: Unscrew and remove the dilution tube applicator. Be careful not to spill or spatter solution from the tube. Collect specimens by inserting the applicator stick into at least 3 different sites of the feces to collect approximately 125 mg of feces . 
      For liquid specimens: Hold the pipette vertically, aspirate fecal specimens, and then transfer 125 µL into the specimen collection tube containing the extraction buffer. 
      3) Place the applicator back into the tube and screw the cap tightly. Be careful not to break the tip of the dilution tube. 
      4) Shake the specimen collection tube vigorously to mix the specimen and the extraction buffer. 
      Specimens prepared in the specimen collection tube may be stored for 6 months at -20°C if not tested within 1 hour after preparation. 
      2. Testing 
      1) Remove the test from its sealed pouch, and place it on a clean, level surface. Label the test with patient or control identification. To obtain a best result, the assay should be performed within one hour. 
      2) Using a piece of tissue paper, break the tip of the dilution tube. Hold the tube vertically and dispense 4 drops of solution into the specimen well (S) of the test device. 
      Avoid trapping air bubbles in the specimen well (S), and do not drop any solution in observation window. 
      As the test begins to work, you will see color move across the membrane. 
      3. Wait for the colored band(s) to appear. The result should be read at 10 minutes. 

    RESULTS


    Norovirus GⅠ

    Positive: * A colored band appears in the control band region (C) and another colored band appears in the T1 band region.

    Norovirus GⅡ Positive: * A colored band appears in the control band region (C) and another colored band appears in the T2 line region.

    Norovirus GⅠand GⅡ Positive: * A colored band appears in the control line region (C) and two other colored bands appear in T1 line region and T2 line region respectively.

    NEGATIVE RESULT:

    One colored band appears in the control band region (C). No band appears in the test band region (T1/T2).

     INVALID RESULT:

     

    Control band fails to appear. Results from any test which has not produced a control band at the specified reading time must be discarded. Please review the procedure and repeat with a new test. If the problem persists, discontinue using the kit immediately and contact your local distributor.

    NOTE:

    The intensity of the color in test region (T1/T2) may vary depending on the concentration of aimed substances present in the specimen. Therefore, any shade of color in the test region

    should be considered positive. Besides, the substances level can not be determined by this qualitative test.

    Insufficient specimen volume, incorrect operation procedure, or performing expired tests are the most likely reasons for control band failure.

     




    The inception of this device springs from a growing need for faster diagnostic methodologies that do not compromise on accuracy. The Norovirus, a leading cause of gastroenteritis worldwide, necessitates prompt identification to prevent outbreaks, especially in sensitive environments like hospitals and schools. The innovative technology embedded within the Norovirus (GI/GII) Rapid Test Device harnesses a unique blend of specificity and sensitivity, ensuring results are not only quick but also precise. By focusing on fecal samples, the test circumvents the challenges of detecting this elusive virus in other specimen types, providing a straightforward and reliable diagnostic approach. At the heart of the device's functionality is its visually interpretable strip, which changes color to indicate the presence of Norovirus GI or GII. This method, refined and optimized through QL Biotech's extensive research and development, simplifies the testing process, eliminating the need for sophisticated laboratory equipment or extensive training. Such simplicity, combined with the device's robust design, makes it an ideal choice for a wide range of settings, from advanced medical facilities to field operations in remote areas. Through the integration of the principles of the Test One Step Dengue Test Kits, this device not only elevates the standard of viral gastroenteritis diagnostics but also paves the way for comprehensive viral detection initiatives, underscoring QL Biotech’s commitment to innovation and public health.

  • Previous:
  • Next: