Rapid Adeno Test: HIV 1&2 Human Immunodeficiency Virus Screening Device

Short Description:

The HIV 1/2 Human Immunodeficiency Virus Rapid Test Device/Strip (Whole Blood/Serum/Plasma) is a rapid chromatographic immunoassay for the qualitative detection of antibodies to HIV 1 and/or HIV 2 in whole blood,serum or plasma.


Product Detail

Product Tags

Introducing the newly developed Adeno Test: the HIV 1&2 Rapid Test Device. This innovative solution is designed to provide reliable, quick results in the detection of HIV 1 & 2 antibodies present in whole blood, serum, or plasma. The test is grounded in the principles of qualitative, membrane-based immunoassay technology, ensuring accurate and efficient analysis. The Adeno Test is not just a product but a breakthrough in medical science. It is a qualitative, membrane-based immunoassay. This means that it employs a highly specific viral antigen to test for the presence of antibodies to HIV 1 and HIV 2. By combining whole blood, serum, or plasma with the test reagents, it identifies the presence of HIV antibodies, thereby confirming the potential presence of the virus.

PRINCIPLE

The HIV 1/2 Human Immunodeficiency Virus Rapid Test Device (Whole Blood/Serum/Plasma) is a qualitative, membrane based immunoassay for the detection of antibodies to HIV 1/2 in whole blood, serum or plasma. The membrane is pre‐coated with recombinant HIV antigens. During testing, the whole blood, serum or plasma specimen reacts with HIV antigen coated particles in the test strip. The mixture then migrates upward on the membrane chromatographically by capillary action and reacts with recombinant HIV antigen on the membrane in the test line region. If the specimen contains antibodies to HIV 1 and/or HIV 2, a colored line will appear in the test line region indicating a positive result. If the specimen does not contain HIV 1 and/or HIV 2 antibodies, a colored line will not appear in the test line region indicating a negative result. To serve as a procedural control, a colored line will always appear in the control line region indicating that proper volume of specimen has been added and membrane wicking has occurred.


Product Detail

Brand: QL

Specimens: : Serum/Plasma

Reading time:10 minutes.

Pack:25 T

STORAGE: 2‐30°C

KIT COMPONENTS(Device)

Test devices

Disposable specimen droppers

Buffer  

Package insert


Brand: QL

Specimens: : Whole Blood/Serum/Plasma

Reading time:Strip:15 minutes. Device:10 minutes

Pack:25 T

STORAGE: 2‐30°C

KIT COMPONENTS(Device)

Test devices

Disposable specimen droppers

Buffer  

Package insert

 

KIT COMPONENTS(Strip)

Test strips

Disposable specimen droppers

Buffer

Package insert

Test cards

 

DIRECTIONS FOR USE

    • HIV 1/2 Rapid Test Strip(Whole Blood/Serum/Plasma)
      Allow the test strip, specimen, buffer, and/or controls to equilibrate to room temperature (15‐30°C) prior to testing. 
      1. Remove the test strip from the foil pouch and use it as soon as possible. Best results will be obtained if the assay is performed within one hour. 
      2. Peel off the tape from the test card, and stick the test strip in the middle of test card with arrows pointing down on the test card. 
      For Serum, Plasma or Venipuncture Whole Blood specimen: Hold the dropper vertically and transfer 2 drops of serum, plasma or venipuncture whole blood (approximately 80 L) to the 
      “Specimen Pad” of the test strip, then add 1 drop of buffer (approximately 40L) and start the timer. 
      For Fingerstick Whole Blood specimen: 
      To use a capillary tube: Fill the capillary tube and transfer approximately 80L of fingerstick whole blood specimen to the “Specimen Pad” of the test strip, then add 1 drop of buffer (approximately 40L) and start the timer. 
      To use hanging drop: Allow 2 hanging drops of fingerstick whole blood specimen to fall into the center of the “Specimen Pad” on the test strip, then add 1 drop of buffer (approximately 
      40L) and start the timer. 
      3. Wait for the colored line(s) to appear. The result should be read at 15 minutes. Do not interpret results after 20 minutes.
      HIV 1/2 Rapid Test Strip(Serum/Plasma)
      HIV 1/2 Rapid Test Device (Whole Blood/Serum/Plasma)
      Allow the test device, specimen, buffer, and/or controls to equilibrate to room temperature 
      (15‐30°C) prior to testing. 
      1. Remove the test device from the foil pouch and use it as soon as possible. Best results will be obtained if the assay is performed within one hour. 
      2. Place the test device on a clean and level surface. Hold the dropper vertically and transfer 2 drops of serum, plasma (approximately 80L) to the specimen well (S) of the sample pad then add 1 drop of buffer (approximately 40L) and start thetimer. 
      3. Wait for the colored line(s) to appear. The result should be read at 10 minutes. Do not interpret results after 20 minutes. 
      HIV 1/2 Rapid Test Device(Serum/Plasma)
      Allow the test device, specimen, buffer, and/or controls to equilibrate to room temperature (15‐30°C) prior to testing. 
      1. Remove the test device from the foil pouch and use it as soon as possible. Best results will be obtained if the assay is performed within one hour. 
      2. Place the test device on a clean and level surface. 
      For Serum, Plasma specimen: Hold the dropper vertically and transfer 2 drops of serum, plasma (approximately 80L) to the specimen well (S) of the test device, then add 1 drop of buffer (approximately 40L) and start the timer. 
      3. Wait for the colored line(s) to appear. The result should be read at 10 minutes. Do not interpret results after 20 minutes.


    INTERPRETATION OF RESULTS

    INVALIDRESULT:

    Control line (C) falls to appear. Insufficient buffer volume or incorrect procedural techniques are the most likely reasons for control line failure.

    Review the procedure and repeat the procedure with a new test device. If the problem persists, discontinue using the test kit immediately and contact your local distributor.

    *NOTE: The intensity of the color in the test line region (T) will vary depending on the concentration of HIV antibodies present in the specimen. Therefore, any shade of colored in the test region (T) should be considered positive.



    Designed with user-friendliness in mind, this Adeno Test is simple to use. It's geared to be a vital tool for healthcare professionals in laboratories, clinics, or any medical institution that requires quick and efficient HIV testing. The beauty of the Adeno Test is its ability to provide expedient results without compromising on accuracy. In conclusion, the HIV 1&2 Rapid Test Device from QL Biotech is more than a testing strip/device, it's a solution of precision, speed, and reliability. Incorporating the Adeno Test into your medical practice will allow for an enhanced ability to detect and diagnose HIV 1&2, offering an improved approach to patient care, and ultimately, better health outcomes. Discover the power and efficiency of the Adeno Test today.