The H. Pylori Ag Rapid Test Strip (Feces) is a rapid chromatographic immunoassay for the qualitative detection of antigens to H. Pylori in feces to aid in the diagnosis of H. Pylori infection.


Product Detail

Product Tags

In the domain of medical diagnostics, the quest for more efficient, accurate, and user-friendly testing solutions is perpetual. At the forefront of this innovation, QL Biotech proudly introduces our cutting-edge H Pylori Test Strips, a rapid test device designed to revolutionize the detection and management of Helicobacter pylori (H. Pylori) infections. H. Pylori, a type of bacteria found in the digestive tract, has been linked to the development of duodenal and gastric ulcers, and in some cases, can lead to stomach cancer. Early detection and treatment are crucial in managing H. Pylori-related complications. This is where QL Biotech steps in, with our commitment to enhancing healthcare outcomes through innovative diagnostic tools. Our H Pylori Test Strips leverage a highly sensitive and specific immunoassay to detect H. Pylori antigens in human feces. This non-invasive testing method is not only convenient but also ensures patient compliance, making it an ideal choice for both healthcare professionals and home users. The simplicity of the procedure, which can be summarized in a few straightforward steps, does not compromise its accuracy. The test can be administered without the need for specialized equipment or extensive training, providing results within minutes.

PRINCIPLE

The H. Pylori Ag Rapid Test Devic/Strip (Feces) is a non‐invasive lateral flow assay, rapid, precise and easy to perform.This test makes use of specific antibodies against H. Pylori antigen adsorbed onto a reactive membrane. If H. Pylori is present in stool specimen, the specific antigen is bound by the second antibody which is conjugated with colloidal gold particles. A generic antibody, fixed onto the reactive membrane, in shape of the band, is able to capture the second conjugated antibody, assuring the correctness of the test performance.


Product Detail

Brand: QL

Specimens: : Feces

Reading time:10 minutes.

Pack:25 T

STORAGE: 2‐30°C

KIT COMPONENTS(Device)

Individually packed Test Devices/Strips

Each test contains a strip with colored conjugates and reactive reagents pre-spreaded at the

corresponding regions

Tubes with buffer

Phosphate buffered saline and preservative, extract the samples

Package insert

For operation instruction

PROCEDURE

    • Bring tests, specimens, buffer and/or controls to room temperature (15-30°C) before use. 
      1. Specimen collection and pre-treatment: 
      1) Unscrew and remove the dilution tube applicator. Be careful not to spill or spatter solution from the tube. Collect specimens by inserting the applicator stick into at least 3 different sites of the feces. 
      2) Place the applicator back into the tube and screw the cap tightly. Be careful not to break the tip of the dilution tube. 
      3) Shake the specimen collection tube vigorously to mix the specimen and the extraction buffer. Specimens prepared in the specimen collection tube may be stored for 6 months at -20°C if 
      not tested within 1 hour after preparation. 
      2. Testing 
      1) Remove the test from its sealed pouch, and place it on a clean, level surface. Label the test with patient or control identification. To obtain a best result, the assay should be performed within one hour. 
      2) Using a piece of tissue paper, remove the tip of the dilution tube. Hold the tube vertically and dispense 3 drops of solution into the sample well of the Test Device. 
      Avoid trapping air bubbles onto the sample pad, and do not drop any solution in observation window. 
      As the test begins to work, you will see color move across the membrane. 
      3. Wait for the colored band(s) to appear. The result should be read at 10 minutes. Do not interpret the result after 20 minutes.

    INTERPRETATION OF RESULTS


    • POSITIVE RESULT: 
      A colored band appears in the control band region (C) and another colored band appears in the T band region 
      NEGATIVE RESULT: 
      One colored band appears in the control band region (C). No band appears in the test band region (T) 
      INVALID RESULT: 
      Control band fails to appear. Results from any test which has not produced a control band at the specified reading time must be discarded. Please review the procedure and repeat with a new test. If the problem persists, discontinue using the kit immediately and contact your local distributor.
      NOTE: 
      1. The intensity of the color in test region (T) may vary depending on the concentration of aimed substances present in the specimen. Therefore, any shade of color in the test region should be considered positive. Besides, the substances level can not be determined by this qualitative test. 
      2. Insufficient specimen volume, incorrect operation procedure, or performing expired tests are the most likely reasons for control band failure.




    The essence of our product lies in its design and functionality, optimized for reliability and ease of use. Each strip is a product of rigorous scientific research, embodying QL Biotech's dedication to quality and innovation. Whether in a clinical setting or the comfort of your home, our H Pylori Test Strips offer a quick, dependable solution for H. Pylori testing. In conclusion, QL Biotech's H Pylori Test Strips represent a significant advancement in the early detection and management of H. Pylori infections. By combining cutting-edge technology with simplicity and accuracy, we strive to improve patient outcomes and advance global health standards. Trust our H Pylori Test Strips for your testing needs and take a step forward in the fight against H. Pylori-related conditions.