Rapid Test HCV Test - QL’s Dengue IgG/IgM Rapid Test Device Innovation

Short Description:

Dengue viruses, a family of four distinct serotypes of viruses (Den 1,2,3,4), are single‐strained, enveloped, positive‐sense RNA viruses.

The test is user friendly, without cumbersome laboratory equipment, and requires minimal staff trainings.


Product Detail

Product Tags

In a fast-paced, rapidly changing world, it's essential to have access to quick, reliable testing methods - something QL biotech understands profoundly. It's with this understanding that we introduce our revolutionary Dengue IgG/IgM Rapid Test HCV Test device. Earning a reputation as a leading innovator in the field of biotechnology, QL Biotech presents an efficient solution for accurate dengue testing. Designed with the user in mind, our device requires only Whole Blood, Serum, or Plasma specimens, thus ensuring the testing process is comfortable and minimally invasive for all involved. To ensure the procedure can be accomplished without unnecessary complexities, our Dengue IgG/IgM Rapid Test HCV Test device delivers your results within a 15-minute timeframe. This stands as a testament to our commitment to efficiency and precision in healthcare solutions.

Product Detail

Brand: QL

Specimens: : Whole Blood/Serum/Plasma

Reading time: in 15 minutes.

Pack:25 T

STORAGE  2‐30°C

KIT COMPONENTS
● Individually packed test devices 

Each device contains a strip with colored  conjugates and reactive reagents pre-spreaded at  the corresponding regions 
● Disposable pipettes 
For adding specimens use 
● Buffer 
Phosphate buffered saline and preservative 
● Package insert 
For operation instruction



1.Dengue IgG/IgM Rapid Test Device Package Insert

PRINCIPLE

The Dengue IgG/IgM Rapid Test Device (Whole Blood/Serum/Plasma) is a qualitative membrane‐based immunoassay for the detection of Dengue antibodies in whole blood, serum, or plasma. This test consists of two components, an IgG component and an IgM component. In the Test region, anti‐human IgM and IgG is coated.

During testing, the specimen reacts with Dengue antigen‐coated particles in the test strip. The mixture then migrates upward on the membrane chromatographically by capillary action and reacts with the anti‐human IgM or IgG in test line region. If the specimen contains IgM or IgG antibodies to Dengue, a colored line will appear in test line region.

Therefore, if the specimen contains Dengue IgM antibodies, a colored line will appear in test line region 1. If the specimen contains Dengue IgG antibodies, a colored line will appear in test line region 2. If the specimen does not contain Dengue antibodies, no colored line will appear in either of the test line regions, indicating a negative result. To serve as a procedural control, a colored line will always appeared in the control line region, indicating that the proper volume of specimen has been added and membrane wicking has occurred.

ASSAY PROCEDURE

Bring the specimen and test components to room temperature Mix the specimen well prior to assay once thawed. Place the test device on a clean, flat surface.

Forwholebloodsample:

Fill the dropper with the specimen then add 1 dropper of specimen into the sample well. The volume is around 10µL. Making sure that there are no air bubbles. Then add 2 drops (about 80 µL) of Sample Diluent immediately into the sample well.

For Plasma/ Serum sample:

Fill the dropper with the specimen not to exceed the specimen line. The volume of the specimen is around 5µL.

Dispense the entire specimen into the center of the sample well making sure that there are no air bubbles. Then add 2 drops (about 80 µL) of Sample Diluent immediately into the sample well.

Note: Practice a few times prior to testing if you are not familiar with the mini dropper. For better  precision,transfer specimenbyapipettecapabletodeliver5µLofvolume.

Set up a timer. Read the result at 15 minutes. Don’t readresultafter30minutes.Toavoidconfusion,discard  thetestdeviceafter interpretingtheresult.

 

2.Dengue NS1 Rapid Test Device Package Insert

The Dengue NS1 Rapid Test Device is a lateral flow chromatographic immunoassay for the qualitative detection of dengue virus antigen (Dengue Ag) in human whole blood, serum or

plasma. It is intended to be used as a screening test and as an aid in the diagnosis of infection with Dengue viruses. Any reactive specimen with the Dengue NS1 Rapid Test Device must be confirmed with alternative testing method(s) and clinical findings

ASSAY PROCEDURE

Step 1: Bring the specimen and test components to room temperature if refrigerated or frozen. Mix the specimen well prior to assay once thawed.

Step 2: When ready to test, open the pouch at the notch and remove device. Place the test device on a clean, flat surface.

Step 3: Be sure to label the device with specimen’s ID number.

Step 4: For whole blood samples:

Fill the dropper with the specimen then add 2 drops (about 80µL) of specimen and 2 drops of buffer into sample well, making sure that there are no air bubbles.

For Plasma/ Serum samples:

Fill the plastic dropper with the specimen. Holding the dropper vertically, dispense 1 drop (about 40µL) of specimen and 2 drops of buffer into the sample well,making sure that there are no air bubbles.

Step 5: Set up a timer.

Step 6: Read the result at 15 minutes.

Don’t read results after 30 minutes. To avoid confusion, discard the test device after

interpreting the result.

 

3.Dengue IgG/IgM/NS1 Combo Rapid Test Device Package Insert

The Dengue IgG/IgM/NS1 Combo Rapid Test Device is a lateral flow chromatographic immunoassay for the qualitative detection of dengue IgG/ IgM and virus antigen (Dengue Ag) in human whole blood, serum or plasma. It is intended to be used as a screening test and as an aid in the diagnosis of infection with Dengue viruses. Any reactive specimen with the Dengue IgG/IgM/NS1 Combo Rapid Test Device must be confirmed with alternative testing method(s) and clinical findings.

ASSAY PROCEDURE

Step 1: Bring the specimen and test components to room temperature if refrigerated or frozen. Mix the specimen well prior to assay once thawed.

Step 2: When ready to test, open the pouch at the notch and remove device. Place the test device on a clean, flat surface.

Step 3: Be sure to label the device with specimen’s ID number.

Step 4: For whole blood samples:

Fill the dropper with the specimen then add 1 drop (about 10µL) of specimen and 2 drops of buffer into the IgG/ IgM sample well and 4 drops of specimen and 2 drops of buffer into NS1 sample well, making sure that there are no air bubbles.

For Plasma/ Serum samples:

Fill the plastic dropper with the specimen. Holding the dropper vertically, dispense 5 µL of specimen and 2 drops of buffer into the IgG/ IgM sample well and 4 drops of specimen and 1 drops of buffer into the NS1 sample well, making sure that there are no air bubbles.

Step 5:Set up a timer.

Step 6: Results can be read in 15 minutes. Positive results can be visible in as short as 1 minute.

Don’t read results after 30 minutes. To avoid confusion, discard the test device after

interpreting the result

INTERPRETATION OF ASSAY RESULT

Dengue NS1

Dengue IgG/IgM

IgG Positive

IgM Positive

IgG and IgM Positive

NEGATIVE RESULT

 

INVALID RESULT

 

Dengue IgG/IgM/NS1

POSITIVE RESULT:

IgG POSITIVE:

IgM POSITIVE:

IgG AND IgM POSITIVE:



Compatibility with three specimen types - Whole Blood, Serum, and Plasma - sets the Rapid Test HCV Test device apart. This flexible functionality speaks volumes of our determination to make biotechnology testing accessible, convenient, and efficient. The rapid result timeframe of 15 minutes only serves to further emphasize this dedication. Overall, the QL Dengue IgG/IgM Rapid Test HCV Test device is designed to streamline your testing experience, promising reliability and certainty in each test. It portrays an impressive example of QL’s continued commitment to revolutionizing biotechnology. Consider entrusting your testing needs to our device - where efficiency meets precision. Discover the QL difference today with the Dengue IgG/IgM Rapid Test HCV Test device.

  • Previous: