HCG One Step Pregnancy Test Device (Serum/Urine)

Short Description:

The hCG One Step Pregnancy Combo Test Device (Urine/Serum) is composed of the glass fiber strips of monoclonal antibody against human chorionic gonadotropin (hCG), anti-mouse IgG solid cellulose nitrate membrane and the bonders of absorptive colloidal gold - monoclonal antibody Against hCG. It adopts the principles of double antibody sandwich method and immunochromatography to test the hCG in the urine and serum.


Product Detail

Product Tags

KIT COMPONENTS

  • ● Test device
    ● Dropper
    ● Package insert

 

SAMPLE COLLECTION AND PREPARATION

Plasma

Have a certified phlebotomist collect whole blood into a purple, blue or green top collection tube (containing EDTA, citrate or heparin, respectively) by veinpuncture.

Separate the plasma by centrifugation.

Carefully withdraw the plasma for testing, or label and store it at 2-8°C for up to two weeks.

Plasma may be frozen at -20°C for up to one year.

Serum

Have a certified phlebotomist collect whole blood into a red top collection tube (containing no anticoagulants) by veinpuncture.

Allow the blood to clot.

Separate the serum by centrifugation.

Carefully withdraw the serum for testing or label and store it at 2-8°C for up to two weeks. Serum may be frozen at -20°C for up to one year.

Urine

Urinate into a clean, dry cup or container.


PERFORMANCE CHARACTERISTICS

Accuracy

A multi-center clinical evaluation was conducted comparing the results obtained using the hCG

One Step Pregnancy Combo Test Device (Urine/Serum) to another commercially available urine membrane hCG test. The study included 150 urine specimens, and both assays identified 72

negative and 78 positive results. The results demonstrated >99% overall accuracy of the hCG One Step Pregnancy Combo Test Device (Urine/Serum) when compared to the other urine membrane hCG test.

Sensiticity and Specifity

The hCG One Step Pregnancy Combo Test Device (Urine/Serum) detects hCG at a concentration of 25mIU/mL or greater. The test has been standardized to the W.H.O. International Standard. The addition of LH (500mIU/mL), FSH (1,000mIU/mL), and TSH (1,000µIU/mL) to negative (0mIU/mL hCG) and positive (25mIU/mL hCG) specimens showed no cross-reactivity.

Interfering Substances

The following potentially interfering substances were added to hCG negative and positive specimens.

Acetaminophen 20 mg/dL     Caffeine       20 mg/dL

Acetylsalicylic Acid 20 mg/dL   Gentisic Acid   20 mg/dL

Ascorbic Acid 20 mg/dL      Glucose        2 g/dL

Atropine 20 mg/dL          Hemoglobin    1 mg/dL

Bilirubin 2 mg/dL

None of the substances at the concentration tested interfered in the assay.